YETUO YTUP15SUV Low-Organic-Matter Ultrapure Water Purifier
Model: YTUP15SUV
Product Overview:
The YETUO YTUP15SUV Organic-Matter Ultrapure Water Purifier has a water production capacity of 15 L/h and adopts a multi-stage purification process of PF+KDF+AC+RO+UV+DI+TF to produce RO reverse osmosis water and UP ultrapure water. UP water has a resistivity of 18.25 MΩ·cm and a conductivity of 0.054 μS/cm, with total organic carbon (TOC) below 3 ppb and heavy metal ions below 0.01 ppb. The RO water salt rejection rate is ≥99%. The system features a built-in 12 L pressure storage tank and an ultrapure water circulation system, together with online water quality monitoring and automatic RO membrane flushing. An optional dual-wavelength UV lamp assembly is available. It is suitable for applications in biology, pharmaceuticals, chemistry, medicine, microelectronics, semiconductors and food industries, including chromatographic analysis, cell culture, protein purification and molecular biology.
- Product Details
- Product manual
YETUO YTUP15SUV Low-Organic-Matter Ultrapure Water Purifier - Product Features
● Intelligent control: The system automatically stops operating when the water supply is interrupted, shuts off the water supply when the machine stops, automatically replenishes the storage tank and stops filling when the tank is full.
● No-water alarm, full-tank alarm and ultrapure water quality alarm. The ultrapure water alarm threshold can be freely configured.
● Built-in 12 L plastic pressure storage tank with two pressure relief valves helps prevent purified water contamination caused by contact with air while saving laboratory space.
● The entire pipeline system adopts quick-connect fittings and comes standard with an external equipment water-supply outlet. External water storage tanks can be added, with various capacities available to meet different requirements.
● Easy filter replacement: The pretreatment and purification columns can be quickly replaced by opening the rear panel of the machine, complying with GLP requirements.
● Fully automatic pressure sensors and microcomputer control enable automatic purified water production.
● Automatic and manual RO membrane anti-scaling flushing programs help significantly extend the service life of the RO membrane.
● Built-in ultrapure water circulation system helps maintain stable ultrapure water quality and provides water on demand.
● A portable conductivity/TDS test pen with a dry-battery design can measure the conductivity and TDS values of tap water and RO water at any time.
● Online resistivity and water temperature monitoring enables real-time monitoring of deionized water and ultrapure water quality.
● NSF-certified pipelines with quick-connect fittings make purification column replacement convenient.
● Imported high-quality RO membrane combines long service life with high water quality and a high salt rejection rate.
● Optional dual-wavelength UV lamp assembly (185 nm and 254 nm) provides effective sterilization, reduces TOC and expands the system's application range.
● Standard 0.22 μm terminal sterilizing filter with an imported PES (polyethersulfone) composite membrane helps ensure sterile water quality.
Detailed Specifications
| Category | Parameter | Specification |
| Basic Information | Product Type | Low-Organic-Matter Type |
| Basic Information | Model | YTUP15SUV |
| Purification System | System Process | PF+KDF+AC+RO+UV+DI+TF |
| Feed Water | Feed-Water Requirements | Municipal tap water (TDS <250 ppm, temperature 5–45°C, pressure 0.02–0.25 MPa, pH 3–10) |
| UP Ultrapure Water Specifications | ||
| UP Water | Resistivity / Conductivity | 18.25 MΩ·cm at 25°C; 0.054 μS/cm at 25°C (<0.1 μS/cm) |
| UP Water | Heavy Metal Ions | <0.01 ppb |
| UP Water | Total Organic Carbon (TOC) | <3 ppb |
| UP Water | Pyrogens / Endotoxins | N/A |
| UP Water | Microorganisms / Bacteria | <0.1 CFU/ml |
| UP Water | Ribonuclease (RNase) | N/A |
| UP Water | Deoxyribonuclease (DNase) | N/A |
| UP Water | Particles (>0.22 μm) | <1/ml |
| RO Reverse Osmosis Water Specifications | ||
| RO Water | TDS Value | RO water TDS (ppm) ≤ Feed-water TDS × 1% (salt rejection rate ≥99%) |
| RO Water | Divalent Ion Rejection Rate | >99% (when using a new RO membrane) |
| RO Water | Organic Matter Rejection Rate | >99% (when molecular weight is >200 Dalton) |
| RO Water | Particle and Bacterial Rejection Rate | >99% |
| Water Production | System Water Production Capacity (25°C) | 15S: 15 L/h; 30S: 30 L/h; 45S: 45 L/h |
| Water Production | Maximum RO Water Production Rate | 2.0 L/min (with built-in pressure tank enabled) |
| Pressure and Water Quality Monitoring | ||
| Monitoring | Inlet Pressure Monitoring | 0–0.7 MPa, 0–100 PSI |
| Monitoring | Outlet Pressure Monitoring | 0–1.6 MPa, 0–220 PSI |
| Monitoring | Pump Inlet Pressure Monitoring | 0–1.6 MPa, 0–220 PSI |
| Monitoring | Digital Water Quality Monitoring | Conductivity/TDS test pen + LCD online monitoring of resistivity, conductivity, TDS and water temperature |
| General Specifications | ||
| Equipment | Dimensions / Weight | 450 × 450 × 500 mm / 45 kg |
| Equipment | Power Supply / Power | 100–240 V, 50–60 Hz / 120 W |
| Configuration | Standard Configuration | Main unit (including 1 set of purification columns), built-in 12 L pressure storage tank, conductivity/TDS test pen and accessory kit |